F-2600 & F-2700 Vortex Flow Meter Manual 02/15 - 0808-7 / 19204 Page 102
TOF, G, f = Factory use only
Sig. Rev = Signal board hardware and rmware revision
Miro Rev = Microprocessor board hardware and rmware revision
AD, R, T, F, PT, V = Factory use only
SPI Err, Rcv, Sent = Factory use only
5.5.2 LEVEL TWO HIDDEN DIAGNOSTICS VALUES
4-20(1) Zero = Analog counts to calibrate zero on analog output 1.
4-20(1) FScale = Analog counts to calibrate full scale on analog output 1.
4-20(2) Zero = Analog counts to calibrate zero on analog output 2.
4-20(2) FScale = Analog counts to calibrate full scale on analog output 2.
4-20(3) Zero = Analog counts to calibrate zero on analog output 3.
4-20(3) FScale = Analog counts to calibrate full scale on analog output 3.
Ext. 4 mA Cal. = Enter 0 for auto calibration or enter factory supplied A/D counts.
Note: You must connect a known 4 mA input if you are going to calibrate the unit.
Ext. 20 mA Cal. = Enter 0 for auto calibration or enter factory supplied A/D counts.
Note: You must connect a known 20 mA input if you are going to calibrate the unit.
External Input = Enter what the external 4-20 mA input represents; i.e. Temperature 1,
Temperature 2, or Pressure. The meter will use this for its internal calculations.
Ext. Full Scale = Enter the full scale units that correlate to the 20 mA point. Note: It
must be in the units for the selected input type such as Deg F, Deg C, PSIA, Bar A, etc.
Ext. Zero Scale = Same as above but for the 4 mA point
Alarm (1) Test = Used as a test to verify that the alarm circuit is functioning. When
Low is selected, the alarm will initiate a low alarm on the output. When High is
selected, it will give a high alarm on the output.
Alarm (2) Test = Used as a test to verify that the alarm circuit is functioning. When
Low is selected, the alarm will initiate a low alarm on the output. When High is
selected, it will give a high alarm on the output.
Alarm (3) Test = Used as a test to verify that the alarm circuit is functioning. When
Low is selected, the alarm will initiate a low alarm on the output. When High is
selected, it will give a high alarm on the output.
Reynolds Corr. = Reynolds number correction for the ow prole. Set to Enable for
F-2700 insertion, and set to Disable for F-2600 inline.
Gain Control = Manual gain control (factory use only). Leave set at 1.
Filter control = Manual lter control. This value can be changed to any number to
force the  value to a constant. A value of zero activates the automatic lter control
which sets  at a level that oats above the f value.
High Pass Filter = Filter setting – Factory use only
Factory Defaults = Reset factory defaults. If you change this to Yes and press ENTER,
all the factory conguration is lost and you must recongure the entire program.
Consult the factory before performing this process, it is required only in very rare
cases.
Meter Type = Insertion or in-line meter
Cong Code = Factory use only
Test Pulse Out = Force totalizer pulse. Set to Yes and press ENTER to send one pulse.
Very useful to test totalizer counting equipment.
Test Scaled Freq = Enter a frequency value in order to test the scaled frequency
output. Return to 0 to stop the test.