4. The vortex coefcient Ck may be incorrectly set. The Ck is a value in the equation used
to determine if a frequency represents a valid vortex signal given the uid density and
signal amplitude. In practice, the Ck value controls the adaptive lter, , setting. During
ow, view the f and values in the rst level of the hidden diagnostics. The value
should be approximately 10-20% higher than the f value. If you raise the Ck setting in
the Calibration Menu, then the value will increase. The is a low pass lter, so by
increasing it or lowering it, you can alter the range of frequencies that the meter will
accept. If the vortex signal is strong, the value will increase to a large number – this is
correct.
Figure 54. Electronics Stack Sensor Connections
Figure 55. Remote Feed Through Board Sensor Connections
VORTEX
TEMPERATURE
PRESSURE
VORTEX
TEMPERATUREPRESSURE